How to Check Chromatography Column Packing

After a column is newly packed or has been in use for a long time, the packing is checked by passing a small amount of a tracer that does not bind to the resin and measuring the height equivalent to a theoretical plate (HETP) and asymmetry (As). This guide summarizes, based on public sources, the test method, the formulas, how to read the results, how to set acceptance criteria and how to keep tracking performance with process data.

Purification & Chromatography · Published 2026.11.23 · Last reviewed 2026.11.23

Why and When to Check Column Packing

In a well-packed column, the tracer passes through evenly and elutes as a narrow, symmetrical peak.[4][3] If the packing density is uneven, the flow velocity differs from place to place, and the peak broadens or leans to one side.[9][10] The column is therefore tested after packing and before use, and the packing is confirmed to be maintained throughout the period the column is in use.[1][3][9][17]

Test Method: Tracer Pulse Injection

  • A small molecule that does not bind to the resin is used as the tracer. Acetone (detected by UV) or sodium chloride (NaCl, detected by conductivity) is commonly used.[1][4]
  • The injection volume is kept small, at no more than a few percent of the column volume, and the exact volume, concentration and flow rate follow the resin manufacturer's instructions.[1][2]
  • Parts outside the column, such as tubing, connections and the detector, also affect peak shape.[9][10]

Calculations: Plate Number, HETP and Asymmetry

  • Plate number N = 5.54 × (Ve ÷ Wh)². Ve is the elution volume of the peak and Wh is the peak width at half height, both expressed in the same units.[1][2][5]
  • HETP = L ÷ N (L is the bed height); the smaller the value, the better the efficiency.[1][8]
  • Using the reduced plate height h = HETP ÷ particle diameter makes it easier to compare columns with different particle sizes and bed heights.[2][7]
  • Asymmetry As = b ÷ a. At 10% of peak height, measure the width in front of the centerline (a) and behind it (b). The closer to 1, the better; a value above 1 indicates a tailing peak and a value below 1 indicates a fronting peak.[1][3][4]
  • The pharmacopoeial symmetry factor for analytical methods is measured at 5% of peak height, so its values are not mixed with or compared to process column asymmetry (10%).[6][1]

How to Read the Results

  • High asymmetry suggests uneven packing density, a void at the column inlet, or channeling, in which liquid concentrates along one path.[3][10][11]
  • Results also vary with sample volume, flow rate, buffer, tubing and temperature. The plate number is therefore used not as an absolute value but as a reference value for comparisons under the same conditions.[2][8][12]
  • In large process columns, peaks often deviate from a Gaussian shape, so calculations that assume a Gaussian distribution can miss changes in packing.[16][18][19]

Setting Acceptance Criteria

  • Acceptance ranges differ by manufacturer and resin. Start from the manufacturer's criteria for the resin in use.[1][3][20]
  • It is recommended to adjust the criteria to the equipment conditions and to set them from the range of historical results accumulated on the same column and equipment.[13][20]
  • Fix the tracer, injection volume, flow rate, buffer and equipment in a standard operating procedure (SOP) so that every test is run under the same conditions.[1][2][14]

Ongoing Tracking with Process Data

  • Running a pulse test every time takes time and is work done outside the process. Transition analysis is therefore used, calculating HETP and asymmetry from the conductivity and UV curves at buffer changeovers during routine operation.[9][15][16]
  • Tracking HETP and asymmetry trends batch by batch allows packing deterioration to be detected early, when HETP rises beyond its control limit.[17][18]

Column Packing Checklist

Item to CheckWhat to Review
When to TestAfter packing and before use, and periodically during use [1][9][17]
TracerAcetone (UV) or NaCl (conductivity), which do not bind to the resin [1][4]
Test ConditionsSmall injection volume, manufacturer's conditions, the same conditions every time [1][2]
CalculationsN = 5.54(Ve/Wh)², HETP = L/N, h = HETP/particle diameter [1][2][7]
Asymmetryb/a at 10% height; whether it is close to 1 [1][4]
CriteriaManufacturer's criteria + range of in-house historical results [13][20]
TrackingBatch-by-batch HETP and As trends, transition analysis [16][17][18]

Sources

  1. Bio-Rad — Nuvia HR-S Media instruction manual, Section 5: Column efficiency test
  2. GE Healthcare (now Cytiva) — Column instructions 71-5000-15 AD, Conditions
  3. Sigma-Aldrich(Merck) — Column Packing and Preparation (protocol)
  4. US Patent US 10,188,964 B2 Chromatography columns
  5. Japanese Pharmacopoeia (PMDA) — General Tests: Chromatography
  6. British Pharmacopoeia 2013, Appendix — Chromatographic Separation Techniques
  7. IUPAC Gold Book — reduced plate height
  8. Shimadzu — Theoretical plate (technical note)
  9. Lightfoot EN, Moscariello J, Rathore AS — Efficiency Measurements for Chromatography Columns, BioPharm International
  10. Stoll DR — But My Peaks Are Not Gaussian! Part 2: Physical Causes of Peak Asymmetry, LCGC International
  11. Morrison RD, Dolan JW — Peak Fronting, Column Life and Column Conditioning, LCGC
  12. Cytiva — Efficiency test of ReadyToProcess columns (application note)
  13. Thermo Fisher Scientific — POROS column packing guide
  14. Rathore AS et al. — Qualification of a Chromatographic Column: Why and How to Do It, BioPharm International
  15. Larson TM, Davis J, Lam H, Cacia J — Use of Process Data To Assess Chromatographic Performance in Production-Scale Protein Purification Columns, Biotechnology Progress 19:485–492 (2003)
  16. Cui Y, Huang Z, Prior J — Using Direct Transition Analysis in Chromatography, BioPharm International 31(1):34–40
  17. Reckamp J — Using Online Transition Analysis to Predict Chromatography Column Failure, Pharmaceutical Technology 44(4):38–41
  18. US Patent US 11,225,516 B2 Transition analysis method for chromatography column qualification
  19. US Patent US 8,568,586 Automated system and method for monitoring chromatography column performance
  20. Cytiva — Column efficiency testing
This article is a reference summary of published regulations, standards and papers, and does not recommend any particular product. When applying it to an individual process, follow the latest original text of the relevant regulation and the review of your quality unit.
Operated by B2BIO Inc., Bioprocess Division · Report content errors to info@b2bio.co.kr

Talk to us about this topic

Tell us the topic of this article and the conditions you are working with now, and we will point you to suitable products and documents.

Talk to us about this topic
↑