What a Virus Removal Filter Does
Virus removal filters retain viruses mainly by size difference (size exclusion).[4][6] They are much tighter than sterilizing-grade filters that remove bacteria, and are designed to let product molecules smaller than about 20 nm pass while retaining larger viruses.[6][11] The reference document for the viral safety of biopharmaceuticals is ICH Q5A(R2); the ICH final version was issued on 1 November 2023, and the U.S. FDA final guidance in January 2024.[1][2]
Large-Virus and Small-Virus Filters
- Depending on the target, filters are divided into large-virus filters (e.g., retroviruses) and small-virus filters (e.g., parvoviruses).[4][7]
- Because parvoviruses are among the smallest of the small viruses, an approach is used in which removal results shown with parvovirus on a small-virus filter are also applied to larger viruses.[3][4]
- To harmonize performance claims that differed from filter to filter, PDA established a rating method using bacteriophages. The small-virus model is PP7, and the large-virus model is PR772.[7][8]
- This rating is a value for comparing filters under standard conditions. Because actual removal capability varies with the product and solution conditions, it is validated with your own product in a scaled-down model of the actual process.[4][5]
Process Conditions Change Virus Removal
ICH Q5A(R2) regards the throughput of process fluid and buffer flush, pressure, and pressure and flow interruptions as conditions that can affect removal capability.[1][4] In studies, small-virus removal decreased in some cases at low pressure or after a process interruption (depressurization), and the extent varied with the filter type, the product, and pH and ionic strength.[4][5] Therefore, if the actual process includes interruptions, they are also included in the validation study.[5][12] The extent of flow decay due to fouling is another condition to examine.[4][6]
Reducing Fouling: Prefilters
- Components similar in size to the filter pores, such as product aggregates, are considered a major cause of virus filter fouling.[6][13]
- For this reason, placing an adsorptive prefilter upstream to capture aggregates and increase throughput is a common approach.[6][11][13]
- Whether to use a prefilter, and which type is suitable, is determined by testing with the actual process fluid.[11][13]
Sizing and Integrity Testing
- Process scale is determined from results obtained by filtering the actual process fluid through a small device, and the same scaled-down model serves as the basis for the validation study.[4][11]
- Integrity tests include the gold particle test, bubble point, and air diffusion test; in production, the non-destructive air diffusion test is widely used.[9][10]
- Integrity test acceptance values are meaningful only when they are linked to virus removal values.[9][10]
- ICH Q5A(R2) calls for defining process controls so that filter replacement and post-use integrity testing can be carried out.[1][3]
Filter Selection Checklist
| Item to Check | What to Review |
|---|---|
| Retention Target | Large-virus or small-virus filter, parvovirus removal data [3][4][7] |
| Rating Data | Bacteriophage rating (PP7, PR772) and test conditions [7][8] |
| Process Conditions | Throughput, flush volume, pressure, presence of interruptions [1][4][5] |
| In-House Validation | Scaled-down model, actual product, process interruptions included [4][5] |
| Fouling | Flow decay, aggregates, need for a prefilter [6][13] |
| Integrity Testing | Test method, link to virus removal values, post-use testing [1][9][10] |
Sources
- ICH — Q5A(R2) Viral Safety Evaluation of Biotechnology Products Derived from Cell Lines of Human or Animal Origin
- U.S. FDA — Guidance for Industry: Q5A(R2) Viral Safety Evaluation …
- Perez-Caballero D, Murray D, Niederlaender C, Brorson K — Viral Safety for Biotechnology Products, Including Viral Vectors: ICH Q5A Revision 2 Brings Updated and More Comprehensive Guidance, BioProcess International
- Gröner A — Virus Retentive Filters – Effective Virus Removal in the Manufacturing Process of Biologicals, Medical Research Archives (2024-12)
- Strauss D et al. — Characterizing the impact of pressure on virus filtration processes and establishing design spaces to ensure effective parvovirus removal, Biotechnology Progress 33(5):1294–1302 (2017)
- Isu S, Qian X, Zydney AL, Wickramasinghe SR — Process- and Product-Related Foulants in Virus Filtration, Bioengineering 9(4):155 (2022)
- PDA — ANSI/PDA Standard 04-2021, Phage Retention Nomenclature Rating for Small- and Large-Virus Retentive Filters
- PDA Journal of Pharmaceutical Science and Technology 62(5):334 (2008) — PDA Virus Filter Task Force rating method study
- Jang H, Saxena L, Kim S, Lee J, Jin Y — Influence of Air-Diffusion Value Shifts on Virus-Clearance Values, BioProcess International (2024-10-15)
- Sekine S, Komuro M, Sohka T, Sato T — Integrity testing of … virus removal filters used in the manufacture of biological products, Biologicals (2015)
- Roederstein S, Thom V — Virus Filtration Using a High-Throughput Parvovirus-Retentive Membrane, BioPharm International (2013)
- Cytiva — Virus filtration for virus removal
- MilliporeSigma — Virus filtration process development (RP5758)